Specifications

Condition
new
Input mode
floating and defibrillation protection
Lead
Lead
Patient leak current
< 10 µA
Input impedance
≥ 2.5 MΩ
Time constant
≥ 3.2 s
Cmrr
> 105 dB
Filter
power frequency (AC 50 / 60 Hz), EMG (25 Hz / 35 Hz (-3 dB)) and baseline drift
Recording mode
thermal printing system
Paper size
210 mm (W) × 20 m (L), high-speed thermal paper
Paper speed
12.5 mm/s, 25 mm/s, 50 mm/s, error: ± 5 %
Gain (sensitivity)
5, 10, 20 mm / mV, accuracy: ± 2 %Standard sensitivity: 10 mm/mV ± 0.2 mm/mV
Record mode
auto, rhythm and manual record
Measurement parameters
HR, P - R interval, P Duration, QRS Duration, T Duration, Q - T interval, Q - Tc, P Axis, QRS Axis, T Axis, R (V5) amplitude, S (V1) amplitude, R (V5) + S (V1) amplitude
Safety class
class I, defibrillation-proof type CF applied part
Polarizing voltage
± 610 mV
Noise level
≤ 12 µVp-p
Sampling frequency for ecg signal input
32 kHz
Sampling frequency for waveform data processing
1 kHz
Sampling accuracy
24 - bit
Minimum detection signal
a 10 Hz, 20 µV (peak - peak value) sinusoidal signal can yield a visual recorded deflection.
Pacemaker detection channel
II
Accuracy of input signal reproduction
± 5 %
Amplitude quantization
≤ 5µV/LSB
Time deviation in each channel
<100 µs
Subcategory
Ecg & eeg & ekg
Listing ID
73245874

Description

Functions
1. Adopt 10.1'' high-resolution color LCD, operate by full-touch screen and function bu ttons, convenient to use.
2. Sync collection for 12-lead ECG, support 12-lead and Cabrera-lead waveform display. Get high-quality ECG waveform via power frequency filter (50 / 60 Hz), baseline filter and EMG filter (25 / 35 Hz) of ECG signal.
3. Display 3 / 6 / 12 - lead ECG waveform, HR value, print mode, sensitivity, paper speed, filter state, clock, battery power, background grid, measured parameter and interpretation, etc. on one screen, and with prompt functions for lead - off, overload and system working state.
4. Power supply: AC / DC (which can adapt AC 50 / 60Hz), with built-in rechargeable lithium battery, charging circuit, perfect over-current and over-voltage protection circuit.

5. In optimal DC state, up to 10-hour standby time, continuous print for more than 3-hour, which meets the requirements of visiting a patient at home and body examination.
6.

Contact Seller for Price